Hospira Plum A+ Infusion System version 13.4 and prior,...
Critical severity
Unreviewed
Published
May 13, 2022
to the GitHub Advisory Database
•
Updated Feb 1, 2023
Description
Published by the National Vulnerability Database
Mar 25, 2019
Published to the GitHub Advisory Database
May 13, 2022
Last updated
Feb 1, 2023
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.